EU-MDR & MR Clinical Trials Portugal

🇵🇹 Submissions in Portugal under MDR

This article provides a structured overview for MR-scientists on how to prepare and where to submit the required documents for a medical-device clinical investigation in Portugal — including studies using non-CE-marked pulse sequences and/or non-CE-marked RF coils.

Authorities to Submit To

In Portugal, “Requests for authorisation of a clinical investigation” must be made to both:

Both approvals are required before starting the investigation.

Key Documents Required

The required documents can be divided in 5 groups:

1. Application Forms

  • MDR Clinical Investigation application to the Competent Authority and Ethics:
    • Submit applications/notifications to INFARMED electronically via RNEC (national portal) (see also RNEC FAQs for device studies).
    • For interventional device studies, CEIC is the competent ethics committee (may designate a local CES); submit via RNEC.
  • When the EUDAMED CI/PS module is required in Portugal, submit in EUDAMED.
  • Cover letter summarising device, investigational components (non-CE sequence/coil), sites, participants, timelines.

2. Clinical Investigation Documentation

3. Safety and Performance Evidence (MR-specific)

  • IEC 60601-2-33 rationale for MR environments (SAR/local SAR limits; simulations and phantom measurements).
  • RF coil dossier (manufacturer ISO 13485 certificate, coil-file, FDTD/Sim4Life B1+/SAR simulations, phantom thermometry, safety report).
  • Sequence safety (vendor statements that scanner safety controls remain active; watchdogs/limits; access controls to prevent accidental use).
  • Best practices (e.g., ISMRM recommendations for experimental RF hardware) referenced in IB/annexes.

4. Regulatory and Legal Documents

  • Insurance certificate covering participants in Portugal.
  • EU Authorised Representative (if sponsor is outside the EU).
  • Data protection: GDPR compliance statement and (where applicable) DPIA.
  • Participant-facing documents should be provided in Portuguese (national language expectation).

5. Ethics Committee Package (CEIC / CES)

  • PIS & ICF (Portuguese language per CEIC / local CES requirements), privacy wording aligned to GDPR; submit via RNEC.
  • Recruitment materials (flyers/posters).
  • Investigator CVs & GCP certificates, site suitability, institutional permissions.

Submission Pathway

  • Competent Authority (INFARMED): submit the full MDR clinical investigation dossier electronically via RNEC (applications/notifications to INFARMED); consult RNEC FAQs. When mandated, use the EUDAMED CI/PS module.
  • Ethics (CEIC / CES): submit the ethics dossier via RNEC; for interventional device studies, CEIC acts as the competent ethics committee (may designate a local CES).

Note: Portugal uses the national portal RNEC for both the CA (INFARMED) and ethics submissions. Coordinate timelines and notify both bodies of substantial modifications.

MR-specific notes for studies using non-CE sequences and/or non-CE RF coils

  • Define the investigational device system clearly (CE-marked scanner/software + investigational coil/sequence package) and state risk class in CIP/IB.
  • Evidence package: B1+ maps, SAR/local-SAR analyses, worst-case duty-cycle temperature estimates, phantom thermometry at hotspot locations (e.g., eye and cortical regions).
  • Operational controls: show that scanner SAR/gradient watchdogs remain active with investigational components; document password-protected access and prevention of accidental use.
  • Standards mapping: cross-reference IB/CIP evidence to IEC 60601-2-33 and your applicable standards list; cite ISMRM RF-hardware safety recommendations.

Relevant References